USP Glycerin Purity Assay on ZB-624
GC Conditions
Column
Zebron ZB-624, GC Cap. Column 30 m x 0.53 mm x 3.00 µm, Ea
Part No
7HK-G005-36
Phase Name
ZB-624
Injection
Split 10:1 @ 220°C
Carrier Gas
Constant Flow Helium, 4.5 mL/min
Detection
Flame Ionization (FID) @ (250°C)
Detection Info
Sample Notes
Concentrations were 2.0 mg/mL Glycerin, 0.050 mg/mL ethylene glycol, 0.050 mg/mL diethylene glycol, and 0.10 mg/mL 2,2,2-trichloroethanol (IS) Resolution = 13.88 between glycerin and diethylene glycol