25-OH Vitamin D2 and D3 Standards from Serum on Fully Porous 5µm C18 after PPT
LC Conditions
Column
Luna 5 µm C18(2) 100 Å, LC Column 50 x 2 mm, Ea
Brand
Luna
Part No
00B-4252-B0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: 0.05 % Formic acid
B: 5 mM Ammonium acetate + 0.1 % Formic acid in methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
92
8
2
5
92
8
3
205
0
100
4
290
0
100
5
360
92
8
Flow Rate
0.90 mL/min
Temperature
35°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 2.0mm ID, 10/Pk Part No.: AJ0-4286 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S100-TBJ
Pretreatment Note
Note: Avoid over drying eluates to avoid loss of analytes and/or internal standard, which will impact both precision and accuracy of data Reconstitute with 100 µL ethyl acetate/acetonitrile (5:95), and vortex mix for 5 seconds Centrifuge for 10 min at 2000g to settle any precipitate Transfer clear liquid to amber-colored autosampler vial and cap
Condition 1
1 mL Acetonitrile
Condition 2
1 mL Acetonitrile/Water (65:35)
Load
Load 1.60 mL Sample diluted in Acetonitrile/Water (50:50)
Wash 1
0.5 mL Acetonitrile/Water (65:35) at 1 mL/min
Wash 2
0.5 mL Acetonitrile/Water (65:35) at 1 mL/min
Elution 1
0.5 mL Acetonitrile at 1 mL/min
Elution 2
0.5 mL Acetonitrile at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
100 µL ethyl acetate/acetonitrile (5:95)
Order Items Used in This Application
- 1. 25-Hydroxy vitamin D2 (25-OH D2)
- 2. 25-Hydroxy vitamin D3-d6 (25-OH D3-d6)
- 3. 25-Hydroxy vitamin D2 (25-OH D2)
Similar Applications