Amphetamines by LC/MS/MS using a Kinetex XB-C18 2.6µm, 50x2.1mm
LC Conditions
Column
Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4496-AN
Phase Name
XB-C18
Elution Type
Gradient
Mobile Phase
A: 5mM Ammonium formate with 0.1% FA
B: Methanol with 0.1% FA
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
1
30
70
3
3
30
70
4
3
30
70
Flow Rate
400.00 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC C18 for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8782 - Holder Part No.: AJ0-9000
Sample Preparation Details
Pretreatment Note
After eluting add 300µL of 0.5 N methanolic hydrochloride to each sample. Samples can then be blown down under a stream of N2 gas at < 40ºC until dry. Reconstitute samples with 1mL of 10% methanol in 0.1% formic acid. Before injecting onto the LC/MS dilute the samples by a factor of 20 to bring the concentration into a suitable range for analysis.
Load
Load Sample
Wash 1
2 mL of 100 mM sodium acetate buffer (pH 5.0)
Wash 2
2 mL methanol
Dry
10 min at 10 in. of Hg
Elution 1
2mLs of (7:2:1) Ethyl Acetate: Isopropanol:NH4OH mixture
Evaporation
Evaporate to dryness
Reconstitute
1 mL 1mL of 10% methanol in 0.1% formic acid
Order Items Used in This Application
- 1. Amphetamine-d11
- 2. Amphetamine
- 3. Methamphetamine-d14
- 4. Methamphetamine
- 5. MDA-d5
- 6. MDA
- 7. MDMA-d5
- 8. MDMA
- 9. MDEA-d5
- 10. MDEA