Ranitidine(Degraded) by HPLC - Luna 3µm C8(2) 100x4.6mm

Ranitidine(Degraded) by HPLC - Luna 3µm C8(2) 100x4.6mm
LC Conditions
Column

Luna 3 µm C8(2) 100 Å, LC Column 100 x 4.6 mm, Ea

Brand

Luna

Part No

00D-4248-E0

Phase Name

C8(2)

Elution Type

Gradient

Mobile Phase

A: 98:2 40mM potassium phospahte pH 7.1:Acetonitrile

B: 78:22 40mM potassium phospahte pH 7.1:Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

100

0

2

10

0

100

3

15

0

100

4

16

100

0

5

20

100

0

Flow Rate

1.50 mL/min

Temperature

35°C

Detection

UV-Vis Abs.-Diode Array (PDA) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 3.0mm ID, 10/Pk Part No.: AJ0-4290 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 3 µm C8(2) 100 Å, LC Column 100 x 4.6 mm, Ea
00D-4248-E0

Luna 3 µm C8(2) 100 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Ranitidine amino alcohol hemifumarate
  • 2. Ranitidine diamine hemifumarate
  • 3. Ranitidine S-oxide
  • 4. Ranitidine N-oxide
  • 5. Ranitidine complex nitroacetamide
  • 6. Ranitidine