Dispersants in seawater on a ZB-Waxplus 30x0.25x0.25 by GCMS
GC Conditions
Column
Zebron ZB-WAXplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Part No
7HG-G013-11
Phase Name
ZB-WAXplus
Injection
Split 10:1 @ 250°C
Carrier Gas
Constant Flow Helium, 1.5 mL/min
Detection
Mass Selective (MSD) @ (250°C)
Detection Info
Sample Notes
Analytes are 1.7ppm in seawater Mass spectrometer is operated in SIM mode: Window 1 (87 (2-butoxyethanol), 89 (IS)), Window 2 (propyleneglycol and IS (31))
Sample Preparation Details
Condition 1
3 mL Methanol at 2 mL/min
Condition 2
3 mL Water at 2 mL/min
Load
Load 1.00 mL Sample diluted in Seawater
Dry
Dry for 5-10min at full vacuum
Elution 1
3 mL 2% Formic acid in methanol at 1 mL/min
Evaporation
Evaporate to dryness
Order Items Used in This Application
- 1. 2-Butoxyethanol-d4
- 2. 2-Butoxyethanol
- 3. Propylene glycol
- 4. Ethylene glycol (EG)
Similar Applications