Fluoroquinolones by LCMS using Kinetex C8 2.6µm, 50 x 2.1 mm
LC Conditions
Column
Kinetex 2.6 µm C8 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4497-AN
Phase Name
C8
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
5
10
90
3
6
10
90
4
6
10
90
Flow Rate
0.40 mL/min
Temperature
40°C
Detection
Mass Spectrometer (MS) @ (550°C)
Detection Info
Sample Notes
Source temp: 550ºC Polarity: Postive Dwell: 25 msec
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 200 mg / 6 mL, Tubes, 30/Pk
Part No
8B-S100-FCH
Condition 1
3 mL Methanol at 2 mL/min
Condition 2
3 mL Water at 2 mL/min
Load
Load 10.00 mL Sample diluted in Phosphate buffer with 2% TCA
Wash 1
2 at 2
Dry
5min @ >10 in Hg
Elution 1
2 mL Methanol/ACN/Acetic acid 49/49/2 at 1 mL/min
Evaporation
Evaporate to dryness
Reconstitute
500 µL 0.1% Formic acid
Order Items Used in This Application
- 1. Ofloxacin
- 2. Norfloxacin
- 3. Norfloxacin-d5
- 4. Ciprofloxacin
- 5. Ciprofloxacin-d8
- 6. Enrofloxacin
- 7. Difloxacin
- 8. Sparfloxacin
Similar Applications
No Similar applications found.