Humic Acids using Kinetex C8 50x2.1 mm on ABI LC-MS 4000
LC Conditions
Column
Kinetex 2.6 µm C8 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4497-AN
Phase Name
C8
Elution Type
Gradient
Mobile Phase
A: 5mM Ammoniium Acetate
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
2
5
95
3
6
5
95
4
6.01
85
15
Flow Rate
0.40 mL/min
Temperature
0°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata PAH, 1.5 g / 6 mL, Tubes, 30/Pk
Part No
8B-S130-7CH
Condition 1
10 mL Dichloromethane at 5 mL/min
Condition 2
20 mL Methanol at 5 mL/min
Condition 3
20 mL MilliQ Water at 5 mL/min
Load
Load 100.00 mL Sample diluted in 75:25 MilliQ Water:Acetonitrile
Wash 1
5 mL 50:50 Methanol:MilliQ Water at 2 mL/min
Dry
15 seconds
Elution 1
6 mL Dichloromethane at 2 mL/min
Evaporation
Evaporate to dryness
Reconstitute
100 µL Dichlromethane
Order Items Used in This Application
- 1. Humic acids
Similar Applications