Hi Temp Stability - Injection on Aeris WIDEPORE after 1050 cycles
LC Conditions
Column
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 50 x 2.1 mm, Ea
Brand
Aeris
Part No
00B-4482-AN
Phase Name
WIDEPORE XB-C18
Elution Type
Gradient
Mobile Phase
A: Water with 0.1% TFA
B: Acetonitrile with 0.1% TFA
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
3
97
3
3
23
35
65
Flow Rate
0.31 mL/min
Temperature
90°C
Detection
UV-Vis Abs.-Diode Array (PDA) @ (22°C)
Detection Info
Sample Notes
Test was run on a 50x4.6 Aeris WIDEPORE 3.6u XB-C18 run at 1.5 mL/min which is not commercially available at this time. Application was modifed to reflect the use of a 2.1 mm column for those that wish to replicat the data
Order Items Used in This Application
- 1. Apomyoglobin
Similar Applications
No Similar applications found.