Phthalates in Foods by LC/MS /MS on Synergi Polar-RP 2.5µm, 100x2.0 mm
LC Conditions
Column
Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea
Brand
Synergi
Part No
00D-4371-B0
Phase Name
Polar-RP
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in water
B: 0.1% Formic acid in 50:50 Acetonitrile: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
60
40
2
1
50
50
3
6
50
50
4
11
15
85
5
14
15
85
6
14.1
60
40
7
19
60
40
Flow Rate
0.40 mL/min
Temperature
22°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC Phenyl for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8788 - Holder Part No.: AJ0-9000
Order Items Used in This Application
- 1. bis(2-Methoxyethyl)phthalate
- 2. Dimethyl phthalate
- 3. bis(2-Ethoxyethyl)phthalate
- 4. Diethyl phthalate
- 5. Diallyl phthalate
- 6. Di-n-propyl phthalate
- 7. Diisobutyl phthalate
- 8. Di-n-butyl phthalate
- 9. bis(2-n-Butoxyethyl)phthalate
- 10. Dibenzylbutyl phthalate
- 11. Dipentyl phthalate
- 12. Dicyclohexyl phthalate
- 13. Di-4-methyl-2-pentyl phthalate
- 14. Dihexyl phthalate
- 15. Diheptyl phthalate
- 16. bis(2-Ethylhexyl)phthalate
- 17. Di-n-octyl phthalate
- 18. Diisononyl phthalate
- 19. Di-n-nonyl phthalate
- 20. Diisodecyl phthalate
Similar Applications