Synthetic Cannabinoids (Parents) by GC-MS on a ZB-Drug-1

Synthetic Cannabinoids (Parents) by GC-MS on a ZB-Drug-1
GC Conditions
Column

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

Part No

7CD-G023-08

Phase Name

ZB-Drug-1

Injection

Split 10:1 @ 260°C

Carrier Gas

Constant Flow Helium, 1.2 mL/min

Detection

Mass Selective (MSD) @ (320°C)

Detection Info
Sample Notes

SIM ions: CP47,497(215*,233,318); CP47,497(215*,233,332); HU-210(302,386,330); JWH-073(327,284,310);JWH-073-D7(334,317); JWH-018(341,284,324); JWH-018-D9(350,333); JWH-200(100,384,56)

Sample Preparation Details
Load

Load 2.00 mL Sample diluted in Urine

Wash 1

2 mL 10% Methanol at 2 mL/min

Dry

5 min at greater than 10 inches mercury

Elution 1

4 mL Ethyl acetate: Isopropanol 85:15 at 1 mL/min

Evaporation

Evaporate to dryness

Reconstitute

100 µL Ethyl acetate

Order Items Used in This Application

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
7CD-G023-08

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

View Product

  • 1. CP47,497
  • 2. CP47,497-C8
  • 3. HU-210
  • 4. JWH-073-D7
  • 5. JWH-018-Pentanoic acid metabolite
  • 6. JWH-018-D9
  • 7. JWH018
  • 8. JWH-122
  • 9. JWH-200
Similar Applications

No Similar applications found.