LC Conditions
Column

Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea

Brand

Synergi

Part No

00D-4371-B0

Phase Name

Polar-RP

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid in DI H2O

B: 0.1% Formic Acid in MeOH

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

25

75

2

2

20

80

3

3.8

20

80

4

3.81

25

75

5

6

25

75

Flow Rate

400.00 mL/min

Temperature

22°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC Phenyl for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8788 - Holder Part No.: AJ0-9000

Order Items Used in This Application

Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea
00D-4371-B0

Synergi 2.5 µm Polar-RP 100 Å, LC Column 100 x 2 mm, Ea

View Product

  • 1. 25 Hydroxy vitamin D3
  • 2. 25-OH-Vit D3-3d
  • 3. 3-Epi-25-Hydroxy vitamin D3
  • 4. 25 Hydroxy vitamin D2
  • 5. impurity (OH-Vit D3)
  • 6. impurity (OH-Vit D2)