Quantification of Acidic Pesticides by QuEchERS and LC/MS/MS

Quantification of Acidic Pesticides by QuEchERS and LC/MS/MS
LC Conditions
Column

Synergi 4 µm Fusion-RP 80 Å, LC Column 50 x 2 mm, Ea

Brand

Synergi

Part No

00B-4424-B0

Phase Name

Fusion-RP

Elution Type

Gradient

Mobile Phase

A: 5mM Ammonium Formate in Water

B: 5mM Ammonium Formate in Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

1.5

60

40

3

8

40

60

4

10

10

90

5

14

10

90

6

15

95

5

Flow Rate

0.25 mL/min

Temperature

25°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Fusion-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7556 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Synergi 4 µm Fusion-RP 80 Å, LC Column 50 x 2 mm, Ea
00B-4424-B0

Synergi 4 µm Fusion-RP 80 Å, LC Column 50 x 2 mm, Ea

View Product

  • 1. Clopyralid
  • 2. Clopyralid
  • 3. Imazapyr
  • 4. Imazapyr
  • 5. Dicamba
  • 6. Dicamba
  • 7. Bentazon
  • 8. Bentazon
  • 9. Fluoxypyr
  • 10. Fluoxypyr
  • 11. 2,4,5-TP
  • 12. 2,4,5-TP
  • 13. Bromoxynil
  • 14. Bromoxynil
  • 15. MCPA
  • 16. MCPA
  • 17. Triclopyr
  • 18. Triclopyr
  • 19. 2,4-DB
  • 20. 2,4-DB
  • 21. Ioxynil
  • 22. Ioxynil
  • 23. 2,4,5–T
  • 24. 2,4,5–T
  • 25. MCPP
  • 26. MCPP
  • 27. 2,4-D
  • 28. 2,4-D
  • 29. 2,4–DP
  • 30. 2,4–DP
  • 31. MCPB
  • 32. MCPB
  • 33. Dinoterb
  • 34. Dinoterb
  • 35. Dinoseb
  • 36. PCP
  • 37. PCP
Similar Applications

No Similar applications found.