Quantification of Acidic Pesticides by QuEchERS and LC/MS/MS
LC Conditions
Column
Synergi 4 µm Fusion-RP 80 Å, LC Column 50 x 2 mm, Ea
Brand
Synergi
Part No
00B-4424-B0
Phase Name
Fusion-RP
Elution Type
Gradient
Mobile Phase
A: 5mM Ammonium Formate in Water
B: 5mM Ammonium Formate in Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
1.5
60
40
3
8
40
60
4
10
10
90
5
14
10
90
6
15
95
5
Flow Rate
0.25 mL/min
Temperature
25°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Fusion-RP 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7556 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Clopyralid
- 2. Clopyralid
- 3. Imazapyr
- 4. Imazapyr
- 5. Dicamba
- 6. Dicamba
- 7. Bentazon
- 8. Bentazon
- 9. Fluoxypyr
- 10. Fluoxypyr
- 11. 2,4,5-TP
- 12. 2,4,5-TP
- 13. Bromoxynil
- 14. Bromoxynil
- 15. MCPA
- 16. MCPA
- 17. Triclopyr
- 18. Triclopyr
- 19. 2,4-DB
- 20. 2,4-DB
- 21. Ioxynil
- 22. Ioxynil
- 23. 2,4,5–T
- 24. 2,4,5–T
- 25. MCPP
- 26. MCPP
- 27. 2,4-D
- 28. 2,4-D
- 29. 2,4–DP
- 30. 2,4–DP
- 31. MCPB
- 32. MCPB
- 33. Dinoterb
- 34. Dinoterb
- 35. Dinoseb
- 36. PCP
- 37. PCP