Quantification of Carbamates Fungicides in Beverage by LC/MS/MS
LC Conditions
Column
Gemini 3 µm C18 110 Å, LC Column 50 x 2 mm, Ea
Brand
Gemini
Part No
00B-4439-B0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 5mM ammonium formate
B: 5mM ammonium formate in acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
2.5
5
95
3
4
5
95
4
4.01
85
15
5
8
85
15
Flow Rate
0.50 mL/min
Temperature
28°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S100-TBJ
Condition 1
1 mL Methanol
Condition 2
1 mL 25 mM Ammonium acetete pH 5.5
Load
Load 1.40 mL Sample diluted in pretreated sample
Wash 1
1 mL 25 mM Ammonium acetete pH 5.5
Wash 2
1 mL Methanol/Water (15:85)
Dry
2 min at 10mm Hg
Elution 1
0.5 mL 5% Formic acid in Methanol
Evaporation
Evaporate to dryness
Reconstitute
1 mL Initial mobile phase
Order Items Used in This Application
- 1. Aldicarb-sulfone
- 2. Methomyl
- 3. 3-Hydroxycarbofuran
- 4. Oxamyl
- 5. Aldicarb
- 6. Carbofuran
- 7. Carbofuran-d3
- 8. Methiocarb
Similar Applications
No Similar applications found.