Chloramphenicol in Seafood and dairy products by LC/MS/MS

Chloramphenicol in Seafood and dairy products by LC/MS/MS
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4462-AN

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

-2

95

5

2

0

95

5

3

2

95

5

4

2.01

5

95

5

4.5

5

95

Flow Rate

0.40 mL/min

Temperature

20°C

Detection

Tandem Mass Spec (MS-MS) @ (450°C)

Detection Info
Sample Notes

SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC C18 for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8782 - Holder Part No.: AJ0-9000

Sample Preparation Details
Description

Strata-X 33 µm Polymeric Reversed Phase, 60 mg / 3 mL, Tubes, 50/Pk

Part No

8B-S100-UBJ

Condition 1

3 mL Methanol at 1 mL/min

Condition 2

3 mL Water at 1 mL/min

Load

Load 10.00 mL Sample

Wash 1

3 mL Water at 1 mL/min

Dry

5 minutes @ 10 in Hg Vaccum

Elution 1

3 mL Ethyl Acetate at 1 mL/min

Evaporation

Evaporate to dryness

Reconstitute

500 µL 0.5 mL Water/Acetonitrile (80:20)

Order Items Used in This Application

Strata-X 33 µm Polymeric Reversed Phase, 60 mg / 3 mL, Tubes, 50/Pk
8B-S100-UBJ

Strata-X 33 µm Polymeric Reversed Phase, 60 mg / 3 mL, Tubes, 50/Pk

View Product

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4462-AN

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. Chloramphenicol
  • 2. Chloramphenicol
  • 3. Chloramphenicol-d5
-OEChem-06211003152D<br><br> 32 32  0     1  0  0  0  0  0999 V2000<br>    6.0010    4.0000    0.0000 Cl  0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690    4.0000    0.0000 Cl  0  0  0  0  0  0  0  0  0  0  0  0<br>    2.5369    1.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0010    1.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0010    2.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.5369   -4.0000    0.0000 O   0  5  0  0  0  0  0  0  0  0  0  0<br>    4.2690   -4.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690    2.0000    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.4030   -3.5000    0.0000 N   0  3  0  0  0  0  0  0  0  0  0  0<br>    4.2690    1.0000    0.0000 C   0  0  1  0  0  0  0  0  0  0  0  0<br>    3.4030    0.5000    0.0000 C   0  0  1  0  0  0  0  0  0  0  0  0<br>    3.4030   -0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1350    0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690   -1.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.5369   -1.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1350    2.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690   -2.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.5369   -2.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.4030   -2.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1350    3.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8059    1.3100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.4030    1.1200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.7365    0.0251    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.5335    0.0251    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    2.3100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8059   -0.6900    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -0.6900    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.6900    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8059   -2.3100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -2.3100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.5380    0.6900    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.1350    4.1200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1 20  1  0  0  0  0<br>  2 20  1  0  0  0  0<br> 11  3  1  1  0  0  0<br>  3 28  1  0  0  0  0<br>  4 13  1  0  0  0  0<br>  4 31  1  0  0  0  0<br>  5 16  2  0  0  0  0<br>  6  9  1  0  0  0  0<br>  7  9  2  0  0  0  0<br> 10  8  1  6  0  0  0<br>  8 16  1  0  0  0  0<br>  8 25  1  0  0  0  0<br>  9 19  1  0  0  0  0<br> 10 11  1  0  0  0  0<br> 10 13  1  0  0  0  0<br> 10 21  1  0  0  0  0<br> 11 12  1  0  0  0  0<br> 11 22  1  0  0  0  0<br> 12 14  2  0  0  0  0<br> 12 15  1  0  0  0  0<br> 13 23  1  0  0  0  0<br> 13 24  1  0  0  0  0<br> 14 17  1  0  0  0  0<br> 14 26  1  0  0  0  0<br> 15 18  2  0  0  0  0<br> 15 27  1  0  0  0  0<br> 16 20  1  0  0  0  0<br> 17 19  2  0  0  0  0<br> 17 29  1  0  0  0  0<br> 18 19  1  0  0  0  0<br> 18 30  1  0  0  0  0<br> 20 32  1  0  0  0  0<br>M  CHG  2   6  -1   9   1<br>M  END
Compound Name

Chloramphenicol

CID

5959

Molecular Formula

C11H12Cl2N2O5

Molecular Weight

323

No. Hydrogen Bond Acceptors

5

No. Hydrogen Bond Donors

3

Similar Applications

Reversed Phase

Chloramphenicol in Seafood and dairy products by LC/MS/MS

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

20026

Reversed Phase

0.1 ng/g Chloramphenicol in Shrimp using Strata-X and Kinetex C18 50 x 2.1 mm by LCMS

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

19453

Reversed Phase

Chloramphenicol in Shrimp Using Strata-X and Kinetex C18, 50 x 2.1 mm by LCMS

Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea

19302