Penicillins in Meat by LC/MS/MS
LC Conditions
Column
Gemini 3 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea
Brand
Gemini
Part No
00F-4439-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Ammonia in water
B: 95% Acetronitrile, 5% water and 0.1% ammonia
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
98
2
2
2
98
2
3
2.42
90
10
4
6
50
50
5
6.1
0
100
6
6.6
0
100
7
6.8
98
2
8
8
98
2
Flow Rate
0.80 mL/min
Temperature
40°C
Detection
Tandem Mass Spec (MS-MS) @ (600°C)
Detection Info
Sample Notes
This separation uses IMethod Kit Part Number KH0-9047 SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 3.0mm ID, 10/Pk Part No.: AJ0-7597 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata C18-E (55 µm, 70 Å), 200 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S001-FBJ
Pretreatment Note
The equilibration time at conditon 0 was 4 minutes before each injection.
Condition 1
4 mL Methanol at 30 µL/min
Condition 2
4 mL Water at 30 µL/min
Load
Load 2.00 mL Sample diluted in Filtrate
Wash 1
4 mL 0.1% Formic Acid in water
Elution 1
2 mL Acetonitrile at 30 µL/min
Elution 2
2 mL Acetonitrile at 30 µL/min
Evaporation
Evaporate to dryness
Reconstitute
200 µL Water
Order Items Used in This Application
- 1. Amoxicillin
- 2. Ampicillin
- 3. Penicillin G
- 4. Penicillin V
- 5. Oxacillin
- 6. Cloxacillin
- 7. Nafcillin
- 8. Dicloxacillin
Similar Applications