Pharma ABN Test 15 Mix on Optimized Agilent 1100 system using Kinetex 2.6u C18 100x4.6mm

Pharma ABN Test 15 Mix on Optimized Agilent 1100 system using Kinetex 2.6u C18 100x4.6mm
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

Brand

Kinetex

Part No

00D-4462-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid in Water

B: 0.1% Formic Acid in Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

0.2

95

5

3

5

5

95

4

6

5

95

5

6.1

95

5

Flow Rate

1.50 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC C18 for 4.6mm ID Columns, 3/Pk Part No.: AJ0-8768 - Holder Part No.: AJ0-9000

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
00D-4462-E0

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Pyridine
  • 2. Acetaminophen
  • 3. Sulfathiazole
  • 4. Pindolol
  • 5. Benzyl alcohol
  • 6. Phenol
  • 7. Acebutolol
  • 8. Chlorpheniramine
  • 9. Triprolidine
  • 10. Prednisolone
  • 11. 3-Methyl-4-nitrobenzoic acid
  • 12. Nortriptyline
  • 13. 2-Hydroxy-5-methylbenzaldehyde
  • 14. Diflunisal
  • 15. Hexanophenone