Underivatized MMA and Succinic Acid in Human Urine on Gemini C18, 3u, 100x3 mm

Underivatized MMA and Succinic Acid in Human Urine on Gemini C18, 3u, 100x3 mm
LC Conditions
Column

Gemini 3 µm C18 110 Å, LC Column 100 x 3 mm, Ea

Brand

Gemini

Part No

00D-4439-Y0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid in DI H2O

B: 0.1% Formic Acid + 10 mM Ammonium Formate in MeOH

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

85

15

2

1.5

5

95

3

2.5

5

95

4

2.51

85

15

5

4.5

85

15

Flow Rate

700.00 mL/min

Temperature

40°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282

Sample Preparation Details
Description

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk

Part No

8B-S038-TAK

Pretreatment Note

This method is for the analysis of underivatized MMA.

Condition 1

1 mL Methanol

Condition 2

1 mL 25 mM Ammonium Formate

Load

Load Sample

Wash 1

0.5 mL Methanol/Water (50:50)

Dry

Dry under high vacuum for 5-10 min

Elution 1

0.6 mL Ammonium hydroxide/Methanol (2:98)

Elution 2

0.6 mL Ammonium hydroxide/Methanol (2:98)

Evaporation

Evaporate to dryness

Reconstitute

200 µL 0.1% Formic acid

Order Items Used in This Application

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk
8B-S038-TAK

Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk

View Product

Gemini 3 µm C18 110 Å, LC Column 100 x 3 mm, Ea
00D-4439-Y0

Gemini 3 µm C18 110 Å, LC Column 100 x 3 mm, Ea

View Product

  • 1. Succinic acid
  • 2. Methylmalonic acid
Similar Applications

No Similar applications found.