Underivatized MMA and Succinic Acid in Human Urine on Gemini C18, 3u, 100x3 mm
LC Conditions
Column
Gemini 3 µm C18 110 Å, LC Column 100 x 3 mm, Ea
Brand
Gemini
Part No
00D-4439-Y0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in DI H2O
B: 0.1% Formic Acid + 10 mM Ammonium Formate in MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
1.5
5
95
3
2.5
5
95
4
2.51
85
15
5
4.5
85
15
Flow Rate
700.00 mL/min
Temperature
40°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S038-TAK
Pretreatment Note
This method is for the analysis of underivatized MMA.
Condition 1
1 mL Methanol
Condition 2
1 mL 25 mM Ammonium Formate
Load
Load Sample
Wash 1
0.5 mL Methanol/Water (50:50)
Dry
Dry under high vacuum for 5-10 min
Elution 1
0.6 mL Ammonium hydroxide/Methanol (2:98)
Elution 2
0.6 mL Ammonium hydroxide/Methanol (2:98)
Evaporation
Evaporate to dryness
Reconstitute
200 µL 0.1% Formic acid
Order Items Used in This Application
- 1. Succinic acid
- 2. Methylmalonic acid
Similar Applications
No Similar applications found.