Extraction and Analysis of Metronidazole in Facial Cleansers

Extraction and Analysis of Metronidazole in Facial Cleansers
LC Conditions
Column

Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4496-AN

Phase Name

XB-C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid

B: MeOH

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

2

60

40

3

3

60

40

4

3.1

95

5

5

5

95

5

Flow Rate

300.00 mL/min

Temperature

25°C

Detection

Mass Spectrometer (MS) @ (650°C)

Detection Info
Sample Notes

SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC C18 for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8782 - Holder Part No.: AJ0-9000

Sample Preparation Details
Description

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

Part No

8B-S029-TBJ

Pretreatment Note

Evaporate to dryness at 50 C. Reconstitute in 200 uL of 10:90 MeOH:0.1% Formic Acid

Condition 1

1 mL of MeOH

Condition 2

1 mL of 0.1N HCl

Load

Load Sample

Wash 1

3 mL of 0.1N HCl

Wash 2

3 mL of MeOH

Wash 3

6 mL of Ethyl Acetate

Dry

5 min under full vacuum

Elution 1

1 mL of 5:95 NH4OH:MeOH (v:v)

Evaporation

Evaporate to dryness

Reconstitute

200 µL 10:90 MeOH:0.1% Formic Acid

Order Items Used in This Application

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
8B-S029-TBJ

Strata-X-C 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk

View Product

Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4496-AN

Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. Metronidazole
  • 2. 13-C2 15-N-Metronidazole
-OEChem-06211003152D<br><br> 21 21  0     0  0  0  0  0  0999 V2000<br>    2.8940   -1.8644    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.2633    1.5835    0.0000 O   0  5  0  0  0  0  0  0  0  0  0  0<br>    5.7281   -0.0638    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7601    0.6356    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.2601    2.1744    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.5202    0.9143    0.0000 N   0  3  0  0  0  0  0  0  0  0  0  0<br>    3.7601   -0.3644    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.9511    1.2234    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5691    1.2234    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8940   -0.8644    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2601    2.1744    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.9143    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.9721   -0.9470    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.3707   -0.2568    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.6820   -0.2818    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.2835   -0.9721    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.6245    2.6760    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.8084    1.5040    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4103    0.7228    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.1916    0.3247    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3571   -2.1744    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1 10  1  0  0  0  0<br>  1 21  1  0  0  0  0<br>  2  6  1  0  0  0  0<br>  3  6  2  0  0  0  0<br>  4  7  1  0  0  0  0<br>  4  8  1  0  0  0  0<br>  4  9  1  0  0  0  0<br>  5  8  2  0  0  0  0<br>  5 11  1  0  0  0  0<br>  6  9  1  0  0  0  0<br>  7 10  1  0  0  0  0<br>  7 13  1  0  0  0  0<br>  7 14  1  0  0  0  0<br>  8 12  1  0  0  0  0<br>  9 11  2  0  0  0  0<br> 10 15  1  0  0  0  0<br> 10 16  1  0  0  0  0<br> 11 17  1  0  0  0  0<br> 12 18  1  0  0  0  0<br> 12 19  1  0  0  0  0<br> 12 20  1  0  0  0  0<br>M  CHG  2   2  -1   6   1<br>M  END
Compound Name

Metronidazole

CID

4173

Molecular Formula

C6H9N3O3

Molecular Weight

171

No. Hydrogen Bond Acceptors

4

No. Hydrogen Bond Donors

1

Similar Applications

Reversed Phase

Extraction and Analysis of Metronidazole in Facial Cleansers

Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea

20631