Histamine by LCMS using Luna 3um HILIC 100x2.1mm
LC Conditions
Column
Luna 3 µm HILIC 200 Å, LC Column 100 x 2 mm, Ea
Brand
Luna
Part No
00D-4449-B0
Phase Name
HILIC
Elution Type
Gradient
Mobile Phase
A: 50:45:5 ACN:DI H2O: 100 mM Amm. Form, pH 3.22
B: 95:5 ACN:100 mM Amm. Formate, pH 3.22
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
17
83
2
1.5
100
0
3
2.5
100
0
4
2.6
17
83
5
4.5
17
83
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-CW 33 µm Polymeric Weak Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S035-TBJ
Condition 1
1 mL Methanol
Condition 2
1 mL DI water
Load
Load Sample
Wash 1
2 mL DI water
Wash 2
2 mL MeOH
Dry
2 min at high vacuum level, ~20-30 in Hg
Elution 1
0.5 mL 87:13 B:A, fortified to contain 5% formic acid
Evaporation
Evaporate to dryness
Reconstitute
0 µL initial mobile phase
Order Items Used in This Application
- 1. 1-Methyl histamine
- 2. N-Methyl histamine
- 3. Histamine
Similar Applications
No Similar applications found.