Steroid panel by LCMS using Kinetex 2.6u C18 100x3.0mm

Steroid panel by LCMS using Kinetex 2.6u C18 100x3.0mm
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea

Brand

Kinetex

Part No

00D-4462-Y0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: Water

B: Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

2

90

10

3

3

60

40

4

7

30

70

5

10

20

80

6

10.2

5

95

7

13

5

95

8

13.2

90

10

9

16

90

10

Flow Rate

0.55 mL/min

Temperature

22°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea
00D-4462-Y0

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea

View Product

  • 1. DHEAS
  • 2. Cortisol
  • 3. Corticosterone
  • 4. 11-DOC (11-Deoxycortisol)
  • 5. Androstenedione
  • 6. Estradiol
  • 7. Testosterone
  • 8. 17-Hydroxyprogesterone
  • 9. DHEA
  • 10. Progesterone
  • 11. 25 Hydroxy vitamin D3
Similar Applications

No Similar applications found.