Cocaine and opiates by GCMS on a ZB-5MS

Cocaine and opiates by GCMS on a ZB-5MS
GC Conditions
Column

Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea

Part No

7FD-G010-08

Phase Name

ZB-5MS

Injection

Splitless @ 250°C

Carrier Gas

Constant Flow Helium, 1.2 mL/min

Detection

Mass Selective (MSD) @ (320°C)

Detection Info
Sample Notes

Sample is derivatized with BSTFA at 70ºC for 30 minutes. Sim mode: Ecgonine methylester (271), Cocaine (182), Cocaethylene (196), Benzoylecgonine (240), Dihydrocodeine (373), Codeine (371), Ethylmorphine (385), Morphine (429), 6-MAM (399)

Order Items Used in This Application

Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
7FD-G010-08

Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea

View Product

  • 1. Ecgonine methyl ester
  • 2. Cocaine
  • 3. Cocaethylene
  • 4. Benzoylecgonine
  • 5. Dihydrocodeine
  • 6. Codeine
  • 7. Ethylmorphine
  • 8. Morphine
  • 9. 6-MAM