OPA Derivitization in Wine of Tyramine and Histamine using Kinetex 2.6u C18

OPA Derivitization in Wine of Tyramine and Histamine using Kinetex 2.6u C18
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

Brand

Kinetex

Part No

00D-4462-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 20mM Sodium Acetate pH 7.2 w/0.03% Sodium Azide (pre-filtered 0.2um)

B: 50:50 Acetonitrile/Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

97

3

2

20

20

80

3

20.5

97

3

4

23

97

3

Flow Rate

1.20 mL/min

Temperature

30°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
00D-4462-E0

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

View Product

  • 1. Histamine
  • 2. Tyramine
-OEChem-06211003142D<br><br> 17 17  0     0  0  0  0  0  0999 V2000<br>    2.8090   -0.4594    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.2633    0.1795    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3090    1.0794    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5691   -0.1807    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.6180    0.1284    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.3122    0.4885    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.3090    1.0794    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.1284    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2780   -0.7281    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.0577   -0.5624    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.6033    1.0359    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8237    0.8702    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8090   -1.0794    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.6734    1.5810    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4103   -0.0632    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.7240    0.5943    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.3922   -0.4270    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1  5  1  0  0  0  0<br>  1  8  1  0  0  0  0<br>  1 13  1  0  0  0  0<br>  2  6  1  0  0  0  0<br>  2 16  1  0  0  0  0<br>  2 17  1  0  0  0  0<br>  3  7  1  0  0  0  0<br>  3  8  2  0  0  0  0<br>  4  5  1  0  0  0  0<br>  4  6  1  0  0  0  0<br>  4  9  1  0  0  0  0<br>  4 10  1  0  0  0  0<br>  5  7  2  0  0  0  0<br>  6 11  1  0  0  0  0<br>  6 12  1  0  0  0  0<br>  7 14  1  0  0  0  0<br>  8 15  1  0  0  0  0<br>M  END
Compound Name

Histamine

CID

774

Molecular Formula

C5H9N3

Molecular Weight

111

No. Hydrogen Bond Acceptors

2

No. Hydrogen Bond Donors

2

Similar Applications

Reversed Phase

OPA Derivitization of Tyramine and Histamine using Kinetex 2.6u C18

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

20873

Reversed Phase

OPA Derivitization of Histadine and Histamine using Kinetex 2.6u C18

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

20872

Reversed Phase

OPA Derivitization in Wine of Histadine and Histamine using Kinetex 2.6u C18

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

20869

Reversed Phase

OPA Derivitization in Wine of Tyramine and Histamine using Kinetex 2.6u C18

Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea

20868