TMP and Thiamine Standard by Impact PPT on Kinetex 5u C18 100x4.6mm
LC Conditions
Column
Kinetex 5 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
Brand
Kinetex
Part No
00D-4601-E0
Phase Name
C18
Elution Type
Gradient
Mobile Phase
A: 25mM Na2HPO4, 10% Methanol, pH 7.0
B: 25mM Na2HPO4, 70% Methanol, pH 7.0
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
0.25
75
25
3
0.75
75
25
4
3
65
35
5
4
40
60
6
4.5
97
3
7
5
97
3
Flow Rate
1.00 mL/min
Temperature
25°C
Detection
Fluorescence (FLUOR) @ (22°C)
Detection Info
Sample Notes
Using Impact Protein Precipitation Plate (Cat Log #CE0-7565) for sample preparation
Sample Preparation Details
Description
Impact Protein Precipitation, 2mL Square Well Filter Plate, 2/Pk
Part No
CE0-7565
Pretreatment Note
Add 400µL of methanol to a well of a Protein Precipitation Plate; Add 100µL plasma to the well. Mix 3 times with pipette tip (or vortex the whole plate briefly); Wait for 5 minutes; Filter through the Protein Precipitation Plate under vacuum (5 Hg) for 5 minutes. Transfer 50 µL to an autosampler vial (or a well of 96-well plate). Add 100µL water and 25 µL of derivatizing reagent*. Cover the vial with a lid. Vortex for 15 seconds. Put the vial into an autosampler. All conditions are at room temperature (RT). It is ready to inject. *Derivatizing reagent: Potassium ferricyanide [K3Fe(CN)6] was prepared as the following: dissolve 6mg of K3Fe(CN)6 in 15mL 15% sodium hydroxide solution. Prepare fresh daily.
Load
Load Sample
Evaporation
Evaporate to dryness
Order Items Used in This Application
- 1. Thiamine monophosphate
- 2. Thiamine
Similar Applications
No Similar applications found.