Paroxetine EP method on Kinetex 5µm C8 250x4.6 mm
LC Conditions
Column
Kinetex 5 µm C8 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Kinetex
Part No
00G-4608-E0
Phase Name
C8
Elution Type
Gradient
Mobile Phase
A: TFA/THF/Water (5:100:900)
B: TFA/THF/ACN (5:100:900)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
30
80
20
3
50
20
80
4
55
20
80
5
60
80
20
6
65
80
20
Flow Rate
1.00 mL/min
Temperature
40°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. Impurity A
- 2. Impurity E
- 3. Paroxetine
- 4. Impurity C
Similar Applications