Extraction of Cannabinoids from Oral Fluid Using Quantisol(OFC Device) on Kinetex Phenyl Hexyl
LC Conditions
Column
Kinetex 2.6 µm Phenyl-Hexyl 100 Å, LC Column 50 x 4.6 mm, Ea
Brand
Kinetex
Part No
00B-4495-E0
Phase Name
Phenyl-Hexyl
Elution Type
Gradient
Mobile Phase
A: 10 mM Ammonium Formate
B: 0.1% Formic Acid in Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
4
0
100
3
6
0
100
Flow Rate
0.60 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-Drug B 33 µm Polymeric Strong Cation, 30 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S128-TBJ
Pretreatment Note
SPE Conditions for Acids Cartridge: 30mg/3 CC of Strata-X Drug B Part No.: 8B-S128-TBJ Condition: 1 mL MeOH Equilibration: 1 mL Acetonitrile:100 mM sodium acetate buffer, pH 5.0 (30/70) Load: Pre-treated samples as below -Place the Swab into the 500 uL oral fluid spiked with analyte mix; wait 1 min -Transfer the Swab in cartridge containing the buffer (supplied by device) -Add 1 mL Acetonitrile:100 mM sodium acetate buffer, pH 5.0 (30/70) -Vortex 30 sec -Break up the cartridge -Centrifuge at 3000rpm for 20 min -Add 1.5 mL 100 mM sodium acetate buffer, pH 5.0 -Transfer contents Wash1: 1 mL 100 mM sodium acetate buffer, pH 5.0 Wash 2: 1 mL Acetonitrile:100 mM sodium acetate buffer, pH 5.0 (30/70) Dry: 5 mins under 10-20 inches of Hg vacuum Elute: 2x0.5 mL Ethyl acetate:Isopropanol (85/15) Add: 30 uL 0.1% HCl/methanol Dry down : To dryness under N2 (at 40-450C ) Reconstitute: In 500 uL of mobile phase (A/B=55/45) spiked with internal standard Note: Conc. of Spiked samples = 80ng/mL for acids
Condition 1
1 mL Methanol
Condition 2
1 mL Acetonitrile: 100 mM sodium acetate buffer, pH 5.0 (30/70)
Load
Load 0.50 mL Sample
Wash 1
1 mL 100 mM sodium acetate buffer, pH 5.0
Wash 2
1 mL Acetonitrile: 100 mM sodium acetate buffer, pH 5.0 (30/70)
Dry
5 min. under vacuum at 10 inches Hg
Elution 1
0.5 mL Ethyl acetate:Isopropanol (85/15)
Elution 2
0.5 mL Ethyl acetate:Isopropanol (85/15)
Evaporation
Evaporate to dryness
Reconstitute
500 µL mobile phase (A/B=55/45)
Order Items Used in This Application
- 1. JWH073(3hb)
- 2. JWH-018(5hp)
- 3. THC-OH
- 4. AM694
- 5. THC-COOH
- 6. JWH-073
- 7. THC
- 8. JWH-018
Similar Applications