Stimulants by LCMS using Kinetex 2.6u Biphenyl 50x3.0mm

Stimulants by LCMS using Kinetex 2.6u Biphenyl 50x3.0mm
LC Conditions
Column

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea

Brand

Kinetex

Part No

00B-4622-Y0

Phase Name

Biphenyl

Elution Type

Gradient

Mobile Phase

A: Water with 0.1% formic acid

B: Methanol with 0.1% formic acid

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

0.5

90

10

3

2

75

25

4

4.5

20

80

5

4.51

15

85

6

5.5

15

85

7

5.51

90

10

8

7

90

10

Flow Rate

0.60 mL/min

Temperature

40°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
00B-4622-Y0

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea

View Product

  • 1. Amphetamine
  • 2. Methamphetamine
  • 3. Phentermine
  • 4. Ritalinic acid
  • 5. Methylphenidate
  • 6. MDMA
  • 7. MDA
  • 8. MDEA