EPA 8151 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)
GC Conditions
Official Method
EPA 8151A
Column
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Part No
7HM-G028-51
Phase Name
ZB-CLPesticides-1
Injection
Splitless @ 250°C
Oven Program
70 °C for 0.5 min to 190 °C @ 25 °C/min for 1 min to 300 °C @ 11 °C/min for 5 min
Carrier Gas
Constant Flow Helium, 36.0 cm/sec
Detection
Electron Capture (ECD) @ (325°C)
Detection Info
Sample Notes
Analytes are 0.1 ug/mL in Hexane
Order Items Used in This Application
- 1. Dalapon methyl ester
- 2. 3,5-Dichlorobenzoic acid methyl ester (SS)
- 3. 4-Nitroanisole
- 4. DCAA methyl ester (SS)
- 5. Dicamba methyl ester
- 6. MCPP methyl ester
- 7. MCPA Methyl ester
- 8. Dichlorprop, methyl ester
- 9. 4,4'-DBOB (IS)
- 10. 2,4-D methyl ester
- 11. Pentachloroanisole
- 12. 2,4,5-TP, methyl ester
- 13. 2,4,5-T methyl ester
- 14. Chloramben, methyl ester
- 15. 2,4-DB methyl ester
- 16. Dinoseb methyl ester
- 17. Bentazon methyl ester
- 18. DCPA methyl ester (Chlorthal-dimethyl)
- 19. Pichloram methyl ester
- 20. Acifluorfen methyl ester
Similar Applications