EPA 504 by GC/ECD on ZB-CLPesticides-2 (Helium Gas)

EPA 504 by GC/ECD on ZB-CLPesticides-2 (Helium Gas)
GC Conditions
Official Method

EPA 504.1

Column

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea

Part No

7HM-G029-11

Phase Name

ZB-CLPesticides-2

Injection

Splitless @ 200°C

Oven Program

30 °C for 2 min to 200 °C @ 30 °C/min

Carrier Gas

Constant Flow Helium, 60.0 cm/sec

Detection

Electron Capture (ECD) @ (220°C)

Detection Info
Sample Notes

2.0 µL pulsed splitless @ 25 psi for 0.5 min

Order Items Used in This Application

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
7HM-G029-11

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea

View Product

  • 1. Bromodichloromethane
  • 2. Chlorodibromomethane
  • 3. 1,2-Dibromomethane (EDB)
  • 4. 1,1,1,2-Tetrachloroethane
  • 5. Bromoform
  • 6. 1,2,3-Trichloropropane
  • 7. 1,2-Dibromo-3-chloropropane (DBCP)