Terlipressin step 2 qc with 0.1% TFA + ACN on knx 5um c18 50x4.6

Terlipressin step 2 qc with 0.1% TFA + ACN on knx 5um c18 50x4.6
LC Conditions
Column

Kinetex 5 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea

Brand

Kinetex

Part No

00B-4601-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% TFA in water

B: acn

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

90

10

2

3

60

40

3

3.1

90

10

4

5

90

10

Flow Rate

2.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 5 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
00B-4601-E0

Kinetex 5 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea

View Product

  • 1. terlipressin
Similar Applications