Mycotoxins extracted from corn using roQ by LC/MS/MS (Negative mode)
LC Conditions
Column
Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4496-AN
Phase Name
XB-C18
Elution Type
Gradient
Mobile Phase
A: 5 mM ammonium acetate w/ 0.5% acetic acid
B: 5 mM ammonium acetate in methanol w/ 0.5% acetic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
2
95
5
3
5
20
80
4
5.2
2
98
5
8
2
98
Flow Rate
0.45 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Sample Notes
roQ extraction: 1) 5 g of corn is extracted using KS0-8909 with 10 mL of water and 10 mL of ACN with 1.0 % formic acid. 2. KSO-8924 is used for dSPE clean up.
Order Items Used in This Application
- 1. Patulin
- 2. Nivalenol
- 3. Deoxynivalenol
- 4. Fusarenon-X
- 5. 3-Acetyldeoxynivalenol
- 6. Zearalenone
- 7. Verruculogen
Similar Applications
No Similar applications found.