Underivatized glyphosate on a Luna 3 µm NH2
LC Conditions
Column
Luna 3 µm NH2 100 Å, LC Column 50 x 2 mm, Ea
Brand
Luna
Part No
00B-4377-B0
Phase Name
NH2
Elution Type
Gradient
Mobile Phase
A: 10 mM Ammonium bicarbonate pH 10
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
20
80
2
5
90
10
3
5.1
20
80
4
8
20
80
Flow Rate
0.40 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. Glyphosate
- 2. 2-Aminoethylphosphonic acid (AMPA)
-OEChem-06211003152D<br><br> 18 17 0 0 0 0 0 0 0999 V2000<br> 3.4030 -0.2015 0.0000 P 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.5369 -0.7015 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.9030 0.6645 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.9030 -1.0675 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 7.7331 0.2985 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 6.8671 -1.2015 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0<br> 5.1350 -0.2015 0.0000 N 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.2690 0.2985 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 6.0010 0.2985 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 6.8671 -0.2015 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0<br> 4.6675 0.7735 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.8705 0.7735 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 5.1350 -0.8215 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 6.3996 0.7735 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 5.6025 0.7735 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 2.0000 -0.3915 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 3.2130 1.2015 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 8.2700 -0.0115 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0<br> 1 2 1 0 0 0 0<br> 1 3 1 0 0 0 0<br> 1 4 2 0 0 0 0<br> 1 8 1 0 0 0 0<br> 2 16 1 0 0 0 0<br> 3 17 1 0 0 0 0<br> 5 10 1 0 0 0 0<br> 5 18 1 0 0 0 0<br> 6 10 2 0 0 0 0<br> 7 8 1 0 0 0 0<br> 7 9 1 0 0 0 0<br> 7 13 1 0 0 0 0<br> 8 11 1 0 0 0 0<br> 8 12 1 0 0 0 0<br> 9 10 1 0 0 0 0<br> 9 14 1 0 0 0 0<br> 9 15 1 0 0 0 0<br>M END
Compound Name
Glyphosate
CID
3496
Molecular Formula
C3H8NO5P
Molecular Weight
169
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
4