Neonicotinoids by LC/MS/MS on a Kinetex Biphenyl
LC Conditions
Column
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Brand
Kinetex
Part No
00B-4622-AN
Phase Name
Biphenyl
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic acid in water
B: 0.1% Formic acid in Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
95
5
2
3
30
70
3
3.1
95
5
4
6
95
5
Flow Rate
0.50 mL/min
Temperature
22°C
Detection
Tandem Mass Spec (MS-MS) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. Dinotefuran
- 2. Nitenpyram
- 3. D3-Thiamethoxam
- 4. Thiamethoxam
- 5. D3-Clothianidin
- 6. Clothianidin
- 7. D4-Imidacloprid
- 8. Imidacloprid
- 9. D3-Acetamiprid
- 10. Acetamiprid
- 11. Thiacloprid
Similar Applications