Terpenes in Marijuana by GC/FID on ZB-35HT
GC Conditions
Column
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Part No
7HG-G025-11
Phase Name
ZB-35HT
Injection
Split 30:1 @ 260°C
Oven Program
80°C for 1 min, 15°/min to 120°, 25°min to 325°C, hold 2.1 min
Carrier Gas
Constant Flow Helium, 2.5 mL/min
Detection
Flame Ionization (FID) @ (325°C)
Detection Info
Sample Notes
FID Gases H2 Flow 30 mL/min Air Flow 300 mL/min Makeup Flow 25 mL/min
Order Items Used in This Application
- 1. Pinene
- 2. Myrcene
- 3. Limonene
- 4. Eucalyptol/Cymene
- 5. Linalool
- 6. Terpineol
- 7. Geraniol
- 8. Pulegone
- 9. Caryophyllene
- 10. pulegon
- 11. Terp 1-1
- 12. Terp 2-1
- 13. caryophyllene
- 14. humulene
- 15. Terp 4-2
- 16. Terp 5-2
- 17. Terp 6-2
- 18. Terp 7-2
Similar Applications