Extraction of Procaine from Serum using Strata-X-C Microelution & a Kinetex Biphenyl HPLC Column

Extraction of Procaine from Serum using Strata-X-C Microelution & a Kinetex Biphenyl HPLC Column
LC Conditions
Column

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea

Brand

Kinetex

Part No

00B-4622-AN

Phase Name

Biphenyl

Elution Type

Gradient

Mobile Phase

A: 0.1% formic acid in water

B: 0.1% formic acid in acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

2

0.5

95

5

3

3

5

95

4

3.5

5

95

5

3.51

95

5

6

5.5

95

5

Flow Rate

0.50 mL/min

Temperature

25°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info
Sample Notes

Strata X-C Elution Plate Protocol Condition 200 µL methanol Equilibrate 200uL water Load 750 µL diluted serum (375 µL plasma diluted 1:1 with 4% phosphoric acid in water) Wash1 200 µL 2% formic acid in water Wash 2 200 µL methanol Elute 25 µL 5% ammonium hydroxide in acetonitrile:methanol(60:40) Inject 1 µL Note: Spiked conc. of analyte was 1 ng/mL serum

Sample Preparation Details
Description

Novum SLE, MINI 96-Well Plate, 1/Pk

Part No

8E-S138-FGA

Pretreatment Note

Sample pre-treatment Dilute 150 uL of human plasma (spiked with 25 ng/mL and 125 ng/mL of cortisone and prednisolone respectively) with 150 uL of 50 mM sodium phosphate dibasic heptahydrate, pH unadjusted. Mix briefly (3-5 sec). Sample loading: Load the sample from pre-treatment step above onto the Novum plate and apply a short and gentle pulse of vacuum (~ 10" of Hg for 20 secs) until the sample has completely entered the media. Wait for 5 minutes. Elution: Dispense 1000 uL of ethyl acetate onto the SLE media and allow the solvent to elute by gravity (~ 5 min elution time) and collect the eluant. Apply vacuum at 5" of Hg for 45 secs to complete the extraction. NOTE: Prolonged application of vacuum will result in elution of plasma out of the SLE media and into the final extracted solvent. Dry down: Evaporate the final extract to complete dryness under slow stream of N2 at 40°C Reconstitute: Reconstitute the dry residue in 150 uL of initial mobile phase fortified with cortisol-D4.

Load

Load 150.00 µL Sample diluted in 150

Elution 1

1000 mL Ethyl acetate

Evaporation

Evaporate to dryness

Reconstitute

150 µL 10mM ammonium acetate in water:methanol(50:50)

Order Items Used in This Application

Novum SLE, MINI 96-Well Plate, 1/Pk
8E-S138-FGA

Novum SLE, MINI 96-Well Plate, 1/Pk

View Product

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
00B-4622-AN

Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea

View Product

  • 1. Procaine
-OEChem-06211003152D<br><br> 37 37  0     0  0  0  0  0  0999 V2000<br>    3.7320    0.4050    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000    0.4050    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981    2.9050    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -4.0950    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981    1.9050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641    3.4050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    3.4050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    1.4050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641    4.4050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    4.4050    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -0.0950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -1.0950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -1.5950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -1.5950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -3.0950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -2.5950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -2.5950    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8101    1.3224    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.2087    2.0127    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.6762    2.8224    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0747    3.5127    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1215    3.5127    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.5200    2.8224    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.5200    1.9876    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1215    1.2973    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0841    4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.4641    5.0250    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.8441    4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.3520    4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    5.0250    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1120    4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690   -1.2850    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4631   -1.2850    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.2690   -2.9050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    1.4631   -2.9050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3291   -4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.4030   -4.4050    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1  8  1  0  0  0  0<br>  1 11  1  0  0  0  0<br>  2 11  2  0  0  0  0<br>  3  5  1  0  0  0  0<br>  3  6  1  0  0  0  0<br>  3  7  1  0  0  0  0<br>  4 15  1  0  0  0  0<br>  4 36  1  0  0  0  0<br>  4 37  1  0  0  0  0<br>  5  8  1  0  0  0  0<br>  5 18  1  0  0  0  0<br>  5 19  1  0  0  0  0<br>  6  9  1  0  0  0  0<br>  6 20  1  0  0  0  0<br>  6 21  1  0  0  0  0<br>  7 10  1  0  0  0  0<br>  7 22  1  0  0  0  0<br>  7 23  1  0  0  0  0<br>  8 24  1  0  0  0  0<br>  8 25  1  0  0  0  0<br>  9 26  1  0  0  0  0<br>  9 27  1  0  0  0  0<br>  9 28  1  0  0  0  0<br> 10 29  1  0  0  0  0<br> 10 30  1  0  0  0  0<br> 10 31  1  0  0  0  0<br> 11 12  1  0  0  0  0<br> 12 13  2  0  0  0  0<br> 12 14  1  0  0  0  0<br> 13 16  1  0  0  0  0<br> 13 32  1  0  0  0  0<br> 14 17  2  0  0  0  0<br> 14 33  1  0  0  0  0<br> 15 16  2  0  0  0  0<br> 15 17  1  0  0  0  0<br> 16 34  1  0  0  0  0<br> 17 35  1  0  0  0  0<br>M  END
Compound Name

Procaine

CID

4914

Molecular Formula

C13H20N2O2

Molecular Weight

236

No. Hydrogen Bond Acceptors

4

No. Hydrogen Bond Donors

1

Similar Applications

Improved Performance

ZB-5MSplus Drugs of Abuse

Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA

23258

Application Specific

Local Anesthetics (Underivatized) on ZB-Drug-1, 10 m x 0.18 mm x 0.18 µm

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

19298

Application Specific

Drug screen by MSD on a ZB-Drug1

Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea

18175

Standard

Common Drug Screen on ZB-50

Zebron ZB-50, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

16012

Standard

Common Drug Screen on ZB-35

Zebron ZB-35, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

16010

Standard

Common Drug Screen on ZB-5MS

Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea

16009

Common Drug Screen on ZB-1MS

16008

Standard

Drugs of Abuse

Zebron ZB-1, GC Cap. Column 15 m x 0.53 mm x 0.50 µm, Ea

1080