Peptide map on Kinetex 2.6 µm EVO C18 (Phosphorylase B)
LC Conditions
Column
Kinetex 2.6 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, EA
Brand
Kinetex
Part No
00F-4725-E0
Phase Name
EVO C18
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA/2% ACN in Water
B: 0.1% TFA/ ACN (10:90)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
98
2
2
68
30
70
Flow Rate
1.50 mL/min
Temperature
60°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
Forced degradation protocol: Reconstitute 1 mg of cefaclor neat standard with 1.0 mL of 1 M HCl. Heat the solution at 60 ºC for 6 hours then neutralize with 1.0 mL of 1 M NaOH. Refrigerate until ready to inject.