Amino Acid Analysis using OPA on Kinetex 5u EVO C18 150x4.6mm
LC Conditions
Column
Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4633-E0
Phase Name
EVO C18
Elution Type
Gradient
Mobile Phase
A: 20mM Potassium Phosphate pH 7.2
B: 50:50 Acetonitrile:Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
97
3
2
20
40
60
3
23
97
3
Flow Rate
1.00 mL/min
Temperature
30°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Order Items Used in This Application
- 1. ASP
- 2. GLU
- 3. SER
- 4. HIS
- 5. THR
- 6. GLY
- 7. ARG
- 8. ALA
- 9. TYR
- 10. MET
- 11. VAL
- 12. CYS
- 13. PHE
- 14. ILE
- 15. LEU
- 16. LYS
Similar Applications