Explosives Mix on Kinetex 5u EVO C18 150x4.6mm

Explosives Mix on Kinetex 5u EVO C18 150x4.6mm
LC Conditions
Column

Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Kinetex

Part No

00F-4633-E0

Phase Name

EVO C18

Elution Type

Isocratic

Mobile Phase

A: Water

B: Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

10

55

45

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4633-E0

Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. HMX
  • 2. RDX
  • 3. 1,3,5-Trinitrobenzene
  • 4. Nitrobenzene
  • 5. 2,4,6-Trinitrotoluene
  • 6. 2-Amino-4,6-dinitrotoluene
  • 7. 2,4-Dinitrotoluene
Similar Applications

No Similar applications found.