Bupropion related compounds on Kinetex 5u EVO C18 150x4.6mm
LC Conditions
Column
Kinetex 5 µm EVO C18 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Kinetex
Part No
00F-4633-E0
Phase Name
EVO C18
Elution Type
Gradient
Mobile Phase
A: 5mM Ammonium Bicarbonate pH 9.0
B: Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
50
50
2
10
20
80
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-CW 33 µm Polymeric Weak Cation, 60 mg / 3 mL, Tubes, 50/Pk
Part No
8B-S035-UBJ
Condition 1
3 mL Metahnol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Load
Load 1.00 mL Sample diluted in 100mM Sodium Phosphate pH 6.8
Wash 1
3 mL 100mM Sodium Phosphate pH 6.8 at 1 mL/min
Dry
5 minutes
Elution 1
2 mL 5% Formic Acid in Acetonitrile at 0.5 mL/min
Evaporation
Evaporate to dryness
Reconstitute
500 µL 50:50 5mM Ammonium Bicarbonate pH 9.0 and Methanol
Order Items Used in This Application
- 1. Hydroxybuproprion
- 2. Erythrobuproprion
- 3. Threobupropion
- 4. Bupropion
Similar Applications