Tumeric pill Analysis on Kinetex 2.6u C18 150x4.6mm

Tumeric pill Analysis on Kinetex 2.6u C18 150x4.6mm
LC Conditions
Column

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Kinetex

Part No

00F-4462-E0

Phase Name

C18

Elution Type

Gradient

Mobile Phase

A: 0.1% Formic Acid in Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

45

55

2

8

30

70

3

9

0

100

4

10

0

100

5

10.1

45

55

6

13

45

55

Flow Rate

1.00 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info

Order Items Used in This Application

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4462-E0

Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Bis-demethoxycurcumin
  • 2. Demethoxycurcumin
  • 3. Curcumin
Similar Applications