ZB-5MSplus Drugs of Abuse

ZB-5MSplus Drugs of Abuse
GC Conditions
Column

Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA

Part No

7HG-G030-11

Phase Name

ZB-5MSplus

Injection

Splitless @ 250°C

Oven Program

70 for 1 min to 150 at 30 ºC/min to 320 at 10 for 5.5 min

Carrier Gas

Constant Flow Helium, 1.4 mL/min

Detection

Mass Spectrometer (MS) @ (250°C)

Detection Info
Sample Notes

MSD: 45 -450 da Sample @ 25ppm

Order Items Used in This Application

Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA
7HG-G030-11

Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA

View Product

  • 1. Nicotine
  • 2. MDA
  • 3. Mdma
  • 4. MBDB
  • 5. Meperidine
  • 6. Benzamphetamine
  • 7. Glutethimide
  • 8. PCP
  • 9. Doxylamine
  • 10. Tramadol
  • 11. Procaine
  • 12. Methadone
  • 13. Topiramate
  • 14. Cocaine
  • 15. Cocaethylene
  • 16. CBD
  • 17. Codeine
  • 18. Ethylmorphine
  • 19. Diazepam
  • 20. Hydrocodone
  • 21. D9-THC
  • 22. 6-mam
  • 23. Oxycodone
Similar Applications