EPA 8151 by GC/ECD on ZB-CLPesticides-2 (Hydrogen Gas)

EPA 8151 by GC/ECD on ZB-CLPesticides-2 (Hydrogen Gas)
GC Conditions
Official Method

EPA 8151

Column

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea

Part No

7HM-G029-11

Phase Name

ZB-CLPesticides-2

Injection

Splitless @ 250°C

Oven Program

70 ºC for 0.5 min to 190 ºC @ 25 ºC/min for 1 min to 300 ºC @ 11 ºC/min for 5 min

Carrier Gas

Constant Flow Hydrogen, 36.0 cm/sec

Detection

Electron Capture (ECD) @ (330°C)

Detection Info
Sample Notes

Analytes are 1000 ng/mL in hexane. Columns are split using a Y-splitter connected to a 5 meter x 0.32 mm ID Z-Guard™ column (7AM-G000-00-GZ0).

Order Items Used in This Application

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
7HM-G029-11

Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea

View Product

  • 1. Dalapon methyl ester
  • 2. 3,5-Dichlorobenzoic acid methyl ester
  • 3. 4-Nitroanisole
  • 4. DCAA methyl ester (SS)
  • 5. Dicamba methyl ester
  • 6. MCPP
  • 7. MCPA
  • 8. Dichlorprop methyl ester
  • 9. 2,4-D methyl ester
  • 10. Pentachloroanisole
  • 11. DBOB (IS)
  • 12. 2,4,5-TP methyl ester
  • 13. Chloramben methyl ester
  • 14. 2,4,5-T methyl ester
  • 15. 2,4-DB methyl ester
  • 16. Bentazon methyl ester
  • 17. Picloram methyl ester
  • 18. Dinoseb methyl ester
  • 19. DCPA methyl ester (Dacthal)
  • 20. Acifluorfen methyl ester
Similar Applications