Chiral Amphetamines with Ephedrines & Pseudoephedrines on Lux 3µm AMP 150x3mm

Chiral Amphetamines with Ephedrines & Pseudoephedrines on Lux 3µm AMP 150x3mm
LC Conditions
Column

Lux 3 µm AMP, LC Column 150 x 3.0 mm, EA

Brand

Lux

Part No

00F-4751-Y0

Phase Name

AMP

Elution Type

Gradient

Mobile Phase

A: 5 mM Ammonium Bicarbonate, adjusted to pH 11 with ammonium hydroxide

B: Methanol

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

40

60

2

10

40

60

3

11

5

95

4

16

5

95

5

16.1

40

60

Flow Rate

0.42 mL/min

Temperature

22°C

Detection

Tandem Mass Spec (MS-MS) @ (22°C)

Detection Info

Order Items Used in This Application

Lux 3 µm AMP, LC Column 150 x 3.0 mm, EA
00F-4751-Y0

Lux 3 µm AMP, LC Column 150 x 3.0 mm, EA

View Product

  • 1. 1S,2R(+)-Ephedrine
  • 2. R,R(-)-Pseudoephedrine
  • 3. S,S(+)-Pseudoephedrine
  • 4. 1R,2S(-)-Ephedrine
  • 5. R(-)-Amphetamine
  • 6. R(-)-Methamphetamine
  • 7. S(+)-Amphetamine
  • 8. S(+)-Methamphetamine
  • 9. Phentermine
  • 10. MDMA
  • 11. MDMA
Similar Applications

No Similar applications found.