FAMEs In Infant Formula by GC/FID on Zebron ZB-FAME
GC Conditions
Column
Zebron ZB-FAME, GC Cap. Column 30m x 0.25mm x 0.2 µm, Ea
Part No
7HG-G033-10
Phase Name
No Phase
Injection
Split 50:1 @ 240°C
Oven Program
100 ºC for 2 min to 140 ºC @ 10 °C/min to 190 ºC @ 3 ºC/min to 260 ºC @ 30 ºC/min for 2 min
Carrier Gas
Constant Flow Helium, 1.2 mL/min
Detection
Flame Ionization (FID) @ (260°C)
Detection Info
Sample Notes
Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Weigh out 500 mg of powdered infant formula into a scintillation vial 2. Dissolve (or dilute) sample in 5 mL toluene 3. Add 6 mL 10 % acetyl chloride solution in methanol 4. Incubate @ 80 °C for 2 hours 5. After incubation add 10 mL 10 % Na2CO3 solution and centrifuge at 5000 rpm for 5 min 6. Extract organic layer for GC analysis
Order Items Used in This Application
- 1. C8:0
- 2. C10:0
- 3. C11:0 (I.S.)
- 4. C12:0
- 5. C14:0
- 6. C16:0
- 7. C18:0
- 8. C18:1 cis 9
- 9. C18:2 cis 9,12
- 10. C18:3 cis 9,12,15
- 11. C20:0
- 12. C20:1 cis 11
- 13. C20:4 cis 5,8,11,14
- 14. C22:0
- 15. C24:0
- 16. C22:6 cis 4,7,10,13,16,19
Similar Applications