Underivatized MMA and Succinic Acid from Plasma on Luna Omega 1.6u PS C18, 50x2.1mm with Strata-X-AW
LC Conditions
Column
Luna Omega 1.6µm PS C18 100Å, LC Column 50 x 2.1 mm, EA
Brand
Luna Omega
Part No
00B-4752-AN
Phase Name
PS C18
Elution Type
Gradient
Mobile Phase
A: 0.1% Formic Acid in H2O
B: 0.1% Formic Acid in ACN
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
98
2
2
2
10
90
3
3
10
90
4
3.01
98
2
5
5
98
2
Flow Rate
0.40 mL/min
Temperature
40°C
Detection
Tandem Mass Spec (MS-MS) @ (600°C)
Detection Info
Sample Notes
MMA (117.0-->72.9; 117.0-->54.9) MMA-D3 (120.1-->75.9)
Sample Preparation Details
Description
Strata-X-AW 33 µm Polymeric Weak Anion, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S038-TAK
Condition 1
1 mL Methanol
Condition 2
1 mL 0.1% Acetic Acid in Water
Load
Load 100.00 µL Sample diluted in 0.1% Acetic Acid in Water
Wash 1
0.5 mL 50:50 Methanol:Water
Dry
5-10 min under high vacuum
Elution 1
0.6 mL 2% Ammonium Hydroxide in Methanol
Elution 2
0.6 mL 2% Ammonium Hydroxide in Methanol
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile Phase A
Order Items Used in This Application
- 1. Interference
- 2. Succinic Acid
- 3. MMA
Similar Applications