FAMEs In Extra Virgin Olive Oil by GC/FID on Zebron ZB-FAME (30 m)
GC Conditions
Column
Zebron ZB-FAME, GC Cap. Column 30m x 0.25mm x 0.2 µm, Ea
Part No
7HG-G033-10
Phase Name
No Phase
Injection
Split 50:1 @ 240°C
Oven Program
100 °C for 2 min to 140 °C @ 10 °C/min to 190 °C @ 3 °C/min to 260 °C @ 30 °C/min for 2 min
Carrier Gas
Constant Flow Helium, 1.2 mL/min
Detection
Flame Ionization (FID) @ (260°C)
Detection Info
Sample Notes
Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Strata® Si-1 Tube, 1 g/6 mL (Part No.: 8B-S012-JCH) on a vacuum or positive pressure manifold 2. Wash cartridge with 6 mL of hexane 3. Load oil solution (0.12 g of oil in 0.5 mL of hexane) 4. Elute with 10 mL of hexane/diethyl ether (87:13) 5. Evaporate eluate under a steady stream of nitrogen 6. Dissolve purified oil residue in 1mL of heptane 7. Add 0.1 mL of 2N potassium hydroxide in methanol to purified oil solution 8. Cap tube and shake vigorously for 15 seconds 9. Leave to separate until upper layer becomes clear 10. Extract upper layer for GC analysis
Sample Preparation Details
Pretreatment Note
To reconstituted sample, add 0.1 mL of 2 N Potassium hydroxide in methanol. Cap tube and shake vigorously for 15 seconds. Leave to separate until upper layer becomes clear. Extract upper layer for GC anlaysis.
Condition 1
6 mL Hexane
Load
Load 0.50 mL Sample diluted in Pretreated sample
Elution 1
10 mL Hexane/Diethyl ether (87:13)
Evaporation
Evaporate to dryness
Reconstitute
1 mL Heptane
Order Items Used in This Application
- 1. C11:0 (I.S.)
- 2. C16:0
- 3. C16:1
- 4. C18:0
- 5. C18:1 cis 9
- 6. C18:2 cis 9,12
- 7. C18:3 cis 9,12,15
- 8. C20:0
- 9. C20:1 cis 11
Similar Applications
No Similar applications found.