Barbiturates from urine by Strata X (60mg/3cc) (TIC)
GC Conditions
Column
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Part No
7CD-G023-08
Phase Name
ZB-Drug-1
Injection
Split 10:1 @ 280°C
Oven Program
145 °C to 300 °C @ 25 °C/min
Carrier Gas
Constant Flow Helium, 0.6 mL/min
Detection
Mass Selective (MSD) @ (320°C)
Detection Info
Sample Notes
Analytes are 300ng/mL in urine derivatized with TMPAH Sim ions: (Quant ion*) D5-Butalbital (*201,184,214): Butalbital (*196,181,138): Amobarbital (*169,184,225): Pentobarbital (*169,184,225): Secobarbital (*196,181,138): D5-Phenobarbital (*233,151,122): Phenobarbital (*232,146,117)
Sample Preparation Details
Pretreatment Note
Reconstitute with 100uL Acetonitrile and 50uL of Thermo Scientific MethElute Reagent (TS49301)
Condition 1
3 mL Methanol at 1 mL/min
Condition 2
3 mL Water at 1 mL/min
Condition 3
1 mL 20mM potassium phosphate pH 2.7 at 1 mL/min
Load
Load 2.00 mL Sample diluted in Diluted Urine
Wash 1
3 mL Water at 1 mL/min
Wash 2
3 mL 100mM acetic acid at 1 mL/min
Dry
Dry column (5 minutes at > 10 inches Hg)
Elution 1
3 mL 85:15 Ethyl Acetate:Isopropyl Alcholol at 0.5 mL/min
Evaporation
Evaporate to dryness
Reconstitute
150 µL Acetonitrile/Thermo Scientific MethElute Reagent
Order Items Used in This Application
- 1. Butalbital-d5
- 2. Butalbital
- 3. Amobarbital
- 4. Pentobarbital
- 5. Secobarbital
- 6. Phenobarbital-d5
- 7. Phenobarbital
Similar Applications