FAMEs In Soybean Oil by GC/FID on Zebron ZB-FAME (60 m)
GC Conditions
Column
Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea
Part No
7KG-G033-10
Phase Name
ZB-FAME
Injection
Split 50:1 @ 240°C
Oven Program
100 °C for 2 min to 165 °C @ 10 °C/min to 200 °C @ 1.5 °C/min to 280 °C @ 15 °C/min for 1 min
Carrier Gas
Constant Flow Helium, 1.2 mL/min
Detection
Flame Ionization (FID) @ (260°C)
Detection Info
Sample Notes
Recommended Accessories: Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Strata® Si-1 Tube, 1 g/6 mL (Part No.: 8B-S012-JCH) on a vacuum or positive pressure manifold 2. Wash cartridge with 6 mL of hexane 3. Load oil solution (0.12 g of oil in 0.5 mL of hexane) 4. Elute with 10 mL of hexane/diethyl ether (87:13) 5. Evaporate eluate under a steady stream of nitrogen 6. Dissolve purified oil residue in 1mL of heptane 7. Add 0.1 mL of 2N potassium hydroxide in methanol to purified oil solution 8. Cap tube and shake vigorously for 15 seconds 9. Leave to separate until upper layer becomes clear 10. Extract upper layer for GC analysis
Order Items Used in This Application
- 1. C11:0
- 2. C16:0
- 3. C18:0
- 4. C18:1 cis 9
- 5. C18:2 cis 9,12
- 6. C18:3 cis 9,12,15
- 7. C20:0
- 8. C20:1 cis 11
- 9. C22:0