FAMEs In Peanut Oil by GC/FID on Zebron ZB-FAME
GC Conditions
Column
Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea
Part No
7KG-G033-10
Phase Name
ZB-FAME
Injection
Split 100:1 @ 240°C
Oven Program
180 °C isothermal
Carrier Gas
Constant Flow Helium, 1.2 mL/min
Detection
Flame Ionization (FID) @ (240°C)
Detection Info
Sample Notes
Recommended Accessories: Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Weigh 0.2 g of sample into a glass test tube 2. Add 3.8 mL of hexane and vortex 3. Condition SPE tube (Strata® Si-1, 2 g/12 mL, Part No.: 8B-S012-KDG) under vacuum with 6 mL of hexane; do not let bed go dry 4. Load sample with light vacuum 5. Elute with 5 mL of hexane:ethyl acetate (87:13) 6. Evaporate under nitrogen at 35 °C 7. Reconstitute with 4 mL of hexane 8. Add 200 µL of 2 M potassium hydroxide/methanol solution and vortex 9. Leave at room temperature for 5 minutes 10. Add 2 mL of deionized water and vortex 11. Leave to separate until upper layer becomes clear 12. Extract upper layer for GC analysis