C18:3 Cis/Trans FAME Isomers by GC/FID on Zebron ZB-FAME (60 m)

C18:3 Cis/Trans FAME Isomers by GC/FID on Zebron ZB-FAME (60 m)
GC Conditions
Column

Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea

Part No

7KG-G033-10

Phase Name

ZB-FAME

Injection

Split 50:1 @ 240°C

Oven Program

160 °C isothermal

Carrier Gas

Constant Flow Helium, 30.0 cm/sec

Detection

Flame Ionization (FID) @ (260°C)

Detection Info
Sample Notes

Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400)

Order Items Used in This Application

Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea
7KG-G033-10

Zebron ZB-FAME, GC Cap. Column 60m x 0.25mm x 0.2 µm, Ea

View Product

  • 1. C18:3 trans 9 trans 12 trans 15
  • 2. C18:3 trans 9 trans 12 cis 15
  • 3. C18:3 trans 9 cis 12 trans 15
  • 4. C18:3 cis 9 cis 12 trans 15
  • 5. C18:3 cis 9 trans 12 trans 15
  • 6. C18:3 cis 9 trans 12 cis 15
  • 7. C18:3 trans 9 cis 12 cis 15
  • 8. C18:3 cis 9 cis 12 cis 15
Similar Applications