Plasma catecholamines using Luna Omega 3um Polar C18 50 x 2.1mm
LC Conditions
Column
Luna Omega 3 µm Polar C18 100 Å, LC Column 50 x 4.6 mm, Ea
Brand
Luna Omega
Part No
00B-4760-E0
Phase Name
Polar C18
Elution Type
Gradient
Mobile Phase
A: 10mM ammonium acetate
B: Acetonitrile with 0.1% formic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
2
3
0
100
Flow Rate
0.40 mL/min
Temperature
22°C
Detection
Electrospray Mass Spec (ESMS) @ (22°C)
Detection Info
Sample Preparation Details
Description
Strata-X-CW Polymeric Weak Cation, microelution 96-well plate, 2mg/well, EA
Part No
8M-S035-4GA
Pretreatment Note
Pool eluents and dilute with 100 µL of 0.1% Formic acid in water
Condition 1
0.2 mL Methanol
Condition 2
0.2 mL 50 mM Ammonium acetate, pH 7
Load
Load 1.00 mL Sample diluted in Pretreated Sample
Wash 1
0.2 mL 50 mM Ammonium acetate, pH 7
Wash 2
0.2 mL Acetonitrile / Isopropyl alcohol (1:1)
Dry
3 minutes under high vacuum
Elution 1
0.025 mL Water / Acetonitrile / Formic acid (85:10:5)
Elution 2
0.025 mL Water / Acetonitrile / Formic acid (85:10:5)
Evaporation
Evaporate to dryness
Order Items Used in This Application
8M-S035-4GA
Strata-X-CW Polymeric Weak Cation, microelution 96-well plate, 2mg/well, EA
- 1. Norepinephrine
- 2. Epinephrine
- 3. Normetanephrine
- 4. Dopamine
- 5. Metanephrine
- 6. Serotonin